From: Evaluation of an in-clinic Serum Amyloid A (SAA) assay and assessment of the effects of storage on SAA samples

Figure 1

Method comparison between in-clinic assay (EVA1) and reference assay (Eiken). Eleven samples with SAA concentrations of 10-270 mg/L (assay range of EVA1) are plotted. The line represents y = x. One sample had a SAA concentration of <10 mg/L with the in-clinic method and 19 mg/L with the reference method (not shown in this figure). Nineteen samples had SAA concentrations of <10 mg/L with both methods and 31 samples had SAA concentrations of >270 mg/L with both methods.