Skip to main content

Table 2 Intra-assay and inter-assay coefficients of variation of samples with high, medium or low cortisol concentrations

From: Validation of an enzyme-linked immunosorbent assay developed for measuring cortisol concentration in human saliva and serum for its applicability to analyze cortisol in pig saliva

Sample concentration

Intra-assay CV

Inter-assay CV

Mean μg/dl

SD

CV %

Mean μg/dl

SD

CV %

High

3.23

0.16

5.0

3.42

0.20

5.9

Medium

2.25

0.04

1.8

1.96

0.29

14.8

Low

0.55

0.01

1.8

0.48

0.06

13.0