Skip to main content

Table 1 Recovery rate in three samples from 3 sows (A, B and C) with different cortisol concentration mixed with accompanying standards

From: Validation of an enzyme-linked immunosorbent assay developed for measuring cortisol concentration in human saliva and serum for its applicability to analyze cortisol in pig saliva

Added concentration1μg/dl

Sow A (0.58 μg/dl)2

Sow B (0.42 μg/dl)2

Sow C (0.54 μg/dl)2

Expected conc.

Detected conc.

Recovery %

Expected conc.

Detected conc.

Recovery %

Expected conc.

Detected conc.

Recovery %

0.03

0.31

0.37

119

0.23

0.21

91

0.29

0.42

145

0.06

0.32

0.34

106

0.24

0.19

79

0.30

0.48

160

0.20

0.39

0.42

108

0.31

0.24

77

0.56

0.56

151

0.60

0.59

0.57

97

0.51

0.49

96

0.57

0.91

160

1.50

1.04

1.31

126

0.96

0.90

94

1.02

1.78

175

4.00

2.29

2.38

104

2.21

2.08

94

2.27

3.00

132

  1. 1Concentration added in 1:1 ratio.
  2. 2The undiluted cortisol concentration of the sample.