Skip to main content

Table 1 Analytical methods used on the ADVIA 2120i

From: Application of the ASVCP guidelines for the establishment of haematologic and biochemical reference intervals in Icelandic horses in Austria

Analyte

Unit

Method

TEo (TEa) (%)

Haematocrit (Hct)

l/l

Calculated from Hct, MCV

3.7 (10)

Erythrocytes (RBC)

T/l

Light scatter

2.1 (10)

Haemoglobin (Hb)

g/l

Cyanide-free colorimetric method

1.9 (10)

MCV

fl

Light scatter/calculated

1.9 (7)

MCH

pg

Calculated

NA

MCHC

g/l

Light scatter/calculated

5.0 (7)

CH

g/l

Light scatter/calculated

NA

CHCM

g/l

Light scatter/calculated

NA

RDW

%

Light scatter/calculated

NA

Platelets (PLT)

G/l

Light scatter

11.1 (25)

MPV

fl

Light scatter

NA

WBC

G/l

MP stain, light scatter

10.2 (15)

MPXI

NU

MP stain, light scatter, calculated

NA

Automated differential

G/l

Light scatter

All within limits

  1. CH cellular haemoglobin, CHCM corpuscular haemoglobin concentration, Hct haematocrit, MCH mean cellular haemoglobin, MCHC mean cellular haemoglobin concentration, MCV: mean cellular volume, MP myeloperoxidase, MPXI myeloperoxidase index, NU no unit, NA not applicable, TEa total allowable error, TEo total observed error, RDW red blood cell distribution width