Skip to main content

Table 2 Multiple reaction monitoring transition ions

From: Quantitative method for analysis of six anticoagulant rodenticides in faeces, applied in a case with repeated samples from a dog

Compound

RT (min)

MRM transitions (m/z)

Cone voltage (V)

Collision energy (eV)

Coumatetralyl

1.86

239.1 > 107.1/91.0

40/30

32/28

Bromadiolone

2.73

511.1 > 251.2/173.0

26

24/42

Difenacoum

2.83

445.3 > 179.1/257.2

30

32/22

Flocoumafen

3.17

543.2 > 159.1/335.2

28

42/24

Brodifacoum

3.27

525.2 > 337.1/178.2

34/55

34/55

Difethialone

3.33

539.1 > 178.1/335.1

36

32/22

Warfarin-d5

1.62

314.2 > 163.1/256.0

24

14/22

  1. Transitions in italics font were used for quantification